Durasphere® Injectable Bulking Agent

Durasphere®  /  Procedure

How the Durasphere® Urethral Bulking Procedure Works

Durasphere® is indicated for the treatment of adult women with stress urinary incontinence (SUI) due to intrinsic sphincter deficiency (ISD), and the urethral bulking procedure is performed cystoscopically using either a transurethral or periurethral approach. This page is a procedural reference for urologists, urogynecologists and gynecologists performing Durasphere injection and covers patient selection, equipment, step-by-step technique for both approaches, post-procedure management, adverse event profiles, and coding.

Durasphere pyrolytic carbon-coated beads beside an injection needle

Durasphere® pyrolytic carbon-coated beads

Patient Selection and Pre-Procedure Workup

Indication

Durasphere is indicated for the treatment of adult women with SUI due to intrinsic sphincter deficiency (ISD). Appropriate candidates are confirmed by urodynamic evaluation; an abdominal leak point pressure below 90 cm H₂O is a key diagnostic criterion for ISD. Federal law restricts this device to sale, distribution, and use by or on the order of a physician trained in diagnostic and therapeutic cystoscopy, and physician training in diagnostic and therapeutic cystoscopy is required prior to use.

Clinicians may reference Durasphere clinical evidence for full study outcomes and post-approval durability data.

Pre-Procedure Workup

Pre-procedure evaluation should confirm ISD as the primary pathophysiologic mechanism of incontinence. Urodynamic studies and cystoscopy are appropriate components of the workup to confirm the diagnosis and assess urethral anatomy.

Contraindications

  • Acute cystitis, urethritis, or other acute genitourinary infection

Equipment and Materials

The Durasphere injection procedure requires Durasphere syringes, Carbon Medical Technologies Injection Needles (CMT catalog numbers 890-XXX), a cystoscope, and related accessories.

Bulking agent

Durasphere EXP

Durasphere is a sterile, nonpyrogenic injectable bulking material composed of pyrolytic carbon-coated graphite beads suspended in a water-based carrier gel containing beta-glucan. The pyrolytic carbon-coated beads are designed to have a minimum dimension of 90 microns and a maximum dimension of 212 microns. For more on how the pyrolytic carbon beads work, see the product technology overview. Durasphere is supplied in individually packaged 1 mL syringes. It does not require refrigeration and should be stored at room temperature (15°C–32°C / 60°F–90°F) prior to use.

Durasphere pre-filled 1 mL syringe
Micrograph of Durasphere pyrolytic carbon-coated beads in carrier gel

Required

Injection needles

Durasphere must be used only with Carbon Medical Technologies Injection Needles (CMT catalog numbers 890-XXX). Use of needles other than those specified may result in clogging of the injection needle.

Required

Cystoscope

For transurethral injection, a 17 to 24 Fr. cystoscope with a minimum working channel of 5 Fr. is required. The InjecSURE® Injection System is compatible with 2.7 mm cystoscopes; see the InjecSURE® compatibility table for specific part number pairings.

Delivery accessory

InjecSURE® Injection System

The InjecSURE® Injection System is a cystoscopic sheath system indicated as an accessory to currently marketed cystoscopes for the delivery of injectable materials into the urethral tissues of the lower urinary tract during cystoscopic procedures. InjecSURE features a non-rotational, dual-port design. The dedicated needle ports are engineered to guide and stabilize the injection needle, supporting needle entry at the desired angle and depth into the urethral submucosa.

The injection system comes with sterile, single-use components: a 20 Ga., 5-inch (127 mm) needle available with either a pencil-point or spinal tip, and a disposable sheath compatible with 2.7 mm cystoscopes.

The InjecSURE Injection System may be used to deliver marketed injectable materials in accordance with their cleared or approved labeling. Injectable materials used in these procedures may include urethral bulking agents, such as Durasphere®.

See how Durasphere compares with other bulking agents to understand delivery system differences across agents.

InjecSURE Injection System sheath and needle

Anesthesia

Anesthesia

The Durasphere injection procedure may be performed under local or regional anesthesia. Anesthesia selection is at the discretion of the treating physician based on patient needs and institutional protocol.

The Durasphere Injection Procedure Step-by-Step

The objective of the procedure is to obtain closure from the bladder neck to the mid-urethra by injecting Durasphere until the tissue surrounding the bladder neck coapts. Durasphere is injected submucosally between the bladder neck and mid-urethra. The injection creates increased tissue bulk and subsequent coaptation from the bladder neck to the mid-urethra.

Durasphere supports both a transurethral and a periurethral approach, the only urethral bulking agent indicated for both.
Cystoscopic view of the urethra and illustration of submucosal bulking

Post-Procedure Care

Immediate post-procedure

It is recommended that patients remain in the clinical setting where they received their Durasphere injection until they are able to void on their own volition. In the event of urinary retention, it can be managed by catheterization in the immediate post-injection phase and with clean intermittent catheterization should it persist.

Retreatment

If incontinence persists after initial injection or if improvement is followed by recurrence of symptoms, treatment may be repeated after sufficient time has passed to evaluate the patient. Retreatment should not occur within 7 days of the previous treatment session. The mean time between treatment sessions in the clinical trial was 5.3 months

Follow-up schedule and activity restrictions

Patients should be evaluated prior to any retreatment session. If incontinence persists or symptoms recur following initial improvement, retreatment may be considered, but no sooner than 7 days after the previous session.

Adverse Events and Management

The adverse event data below are drawn from the Durasphere clinical trial, which involved 300 Durasphere treatment injections in 178 patients. There were no deaths among patients injected with Durasphere during the clinical trial. A total of 87.6% of all adverse events were classified as mild, 11.6% as moderate, and 0.8% as severe. The three events classified as severe (chest pain, renal failure, and myocardial infarction) were reported as unrelated to the device or procedure.

  • Mild 87.6%
  • Moderate 11.6%
  • Severe 0.8%
24.7%

of patients

Urinary urgency

Urinary urgency was reported in 24.7% of patients in the clinical trial (treatment-related urgency in 13%). The majority of urgency events resolved by study end.

16.9%

of patients

Acute urinary retention

Acute retention (duration ≤ 7 days) occurred in 16.9% of patients in the clinical trial; treatment-related acute retention was reported in 16% of patients (29/178). All cases resolved within 7 days via intermittent self-catheterization.

18.0%

of patients

Dysuria

Dysuria was reported in 18.0% of patients overall and in 12% on a treatment-related basis.

29.8%

of patients

Urinary tract infection

UTI was reported in 29.8% of patients overall; treatment-related UTI occurred in 9% of patients. The mean onset of treatment-related UTI was 11 days post-injection compared with 126 days for non-treatment-related events.

33%

transient, first 24 hours

Hematuria

Hematuria was reported transiently in 33% of patients in the 24 hours immediately following injection, with spontaneous resolution. Hematuria beyond 24 hours was reported in 6.7% of patients overall (treatment-related: 6%).

1.7%

of patients

Allergic reaction

Unlike bovine collagen (Contigen), Durasphere does not require pre-treatment intradermal skin testing. Allergic reaction to the antibiotic was reported in 1.7% of patients overall (treatment-related: <1%) in the clinical trial.

Warning

 

Vascular injection

Do not inject Durasphere into blood vessels. Intravascular injection may cause vascular occlusion, platelet aggregation, infarction, or embolic phenomena.

For the full adverse event dataset and outcomes data, see the full safety and outcomes data report.

CPT Coding for Cystoscopy with Durasphere Injection

CPT

51715

Endoscopic injection of implant material into the submucosal tissues of the urethra and/or bladder neck

Cystoscopy with Durasphere injection bills under CPT 51715, which covers endoscopic injection of implant material (natural or synthetic) into the submucosal tissues of the urethra and/or bladder neck for urinary incontinence. This applies to both the transurethral and periurethral approaches.

HCPCS

L8606

Injectable bulking agent, synthetic implant, urinary tract, 1 mL syringe

Durasphere is billed per 1 mL syringe under HCPCS code L8606, “Injectable bulking agent, synthetic implant, urinary tract, 1 ml syringe, includes shipping and necessary supplies.” L8606 is the applicable HCPCS code for currently approved synthetic urethral bulking agents, including Durasphere.

Frequently Asked Questions

CPT 51715 is used for endoscopic injection of implant material into the submucosal tissues of the urethra and/or bladder neck. This code applies to the Durasphere procedure regardless of approach (transurethral or periurethral). The Durasphere product is billed separately under HCPCS L8606 per 1 mL syringe. CPT codes are updated annually; practices should verify the current code with the AMA CPT manual and applicable payer policies prior to billing.

Durasphere is injected using consistent, moderate thumb pressure on the plunger. A high-pressure injection gun is not required and is not appropriate. The resistance encountered during injection provides real-time tactile feedback: if the needle has been placed in muscle rather than submucosal tissue, Durasphere beads will not flow because muscle is too dense. Only the carrier gel will flow under extreme force, and the beads will clog the needle. This resistance response confirms incorrect placement. Correct submucosal placement produces a visible bleb or rise in the mucosal lining at the injection site.

Procedures typically require 4 to 6 mL. Most transurethral and periurethral procedures require 4 or more mL, with a mean initial injection volume of 4.83 mL and a mean total volume including retreatments of 7.55 mL. Individual volume will vary based on urethral anatomy and coaptation endpoint.

The Durasphere injection procedure may be performed under local or regional anesthesia, at the discretion of the treating physician. Specific anesthesia selection will depend on patient factors and institutional protocol.

Follow-up should be scheduled to evaluate treatment response prior to any retreatment decision. If retreatment is needed, it should not occur sooner than 7 days after the previous session.

Brief Summary Statement

Indications: Durasphere Injectable Bulking Agent is indicated for use in the treatment of adult women with stress urinary incontinence (SUI) due to intrinsic sphincteric deficiency (ISD).

Contraindications: The Durasphere agent must not be used in patients with acute cystitis, urethritis, or other acute genitourinary infections.

Warnings/Precautions: Do not inject the Durasphere agent into blood vessels. This may cause vascular occlusion, platelet aggregation, infarction, or embolic phenomena.

Adverse Events: Adverse events related to the Durasphere agent that were reported by >5% of patients during clinical study include: retention, dysuria, urinary urgency, urinary tract infection, hematuria.

Caution: Federal law (U.S.A.) restricts this device to sale, distribution, and use by or on the order of a physician trained in diagnostic and therapeutic cystoscopy.