Durasphere® is indicated for the treatment of adult women with stress urinary incontinence (SUI) due to intrinsic sphincter deficiency (ISD), and the urethral bulking procedure is performed cystoscopically using either a transurethral or periurethral approach. This page is a procedural reference for urologists, urogynecologists and gynecologists performing Durasphere injection and covers patient selection, equipment, step-by-step technique for both approaches, post-procedure management, adverse event profiles, and coding.

Durasphere® pyrolytic carbon-coated beads
Patient Selection and Pre-Procedure Workup
Indication
Durasphere is indicated for the treatment of adult women with SUI due to intrinsic sphincter deficiency (ISD). Appropriate candidates are confirmed by urodynamic evaluation; an abdominal leak point pressure below 90 cm H₂O is a key diagnostic criterion for ISD. Federal law restricts this device to sale, distribution, and use by or on the order of a physician trained in diagnostic and therapeutic cystoscopy, and physician training in diagnostic and therapeutic cystoscopy is required prior to use.
Clinicians may reference Durasphere clinical evidence for full study outcomes and post-approval durability data.
Pre-Procedure Workup
Pre-procedure evaluation should confirm ISD as the primary pathophysiologic mechanism of incontinence. Urodynamic studies and cystoscopy are appropriate components of the workup to confirm the diagnosis and assess urethral anatomy.
Equipment and Materials
The Durasphere injection procedure requires Durasphere syringes, Carbon Medical Technologies Injection Needles (CMT catalog numbers 890-XXX), a cystoscope, and related accessories.
The Durasphere Injection Procedure Step-by-Step
The objective of the procedure is to obtain closure from the bladder neck to the mid-urethra by injecting Durasphere until the tissue surrounding the bladder neck coapts. Durasphere is injected submucosally between the bladder neck and mid-urethra. The injection creates increased tissue bulk and subsequent coaptation from the bladder neck to the mid-urethra.

Transurethral Injection
- 1
Using standard procedure, prepare the patient for cystoscopy. Connect the Durasphere syringe to the CMT Injection Needle and prime the needle. Insert the primed needle into the cystoscope working channel. Insert the scope into the urethra.
- 2
At mid-urethra, choose a position between 4 and 8 o’clock.
- 3
Position the needle bevel toward the urethral lumen. Puncture tissue at a 45° angle until the bevel of the needle is covered with tissue. Do not insert past the bevel.
- 4
Re-angle the scope back parallel to the urethra. Using the circumferential markings on the needle as a guide, tunnel the needle tip toward the bladder neck 1–2 cm, depending on the length between the mid-urethra and bladder neck. After tunneling, the tip of the needle should lie at the proximal urethra.
- 5
Begin injecting Durasphere using consistent, moderate thumb pressure on the plunger. The flow should be even and smooth, similar to priming the needle.
- 6
The submucosal lining should begin to rise at the injection site. By rotating the bevel up and down, the flow of Durasphere can be directed. Continue injecting into this site until the bulge has crossed the midline of the urethra.
- 7
If necessary, choose an opposing site for injection. Repeat steps 3 through 7 at a second injection site. Most procedures will require 4 or more syringes of Durasphere.
- 8
Continue injecting until the tissue surrounding the bladder neck coapts. The bladder neck should be closed when viewing with cystoscopic irrigation on.
Note on tissue plane: If the injection needle is inserted into muscle rather than submucosal tissue, Durasphere beads will not flow because muscle is too dense to accept them. The gel component may flow into the muscle under extreme force, at which point the beads will clog the injection needle. No appreciable bulking will occur with incorrect tissue plane placement.
Transurethral Injection Utilizing InjecSURE®
- 1
Using standard procedure, prepare the patient for cystoscopy.
- 2
Attach InjecSURE to the compatible cystoscope according to the device instructions for use.
- 3
Connect the Durasphere syringe to the CMT Injection Needle and prime the needle.
- 4
Insert the primed needle through one of the InjecSURE needle ports.
- 5
Insert the cystoscope with the attached InjecSURE sheath into the urethra.
- 6
At the mid-urethra, select either the 3 or 9 o’clock position for injection.
- 7
Position the needle bevel toward the urethral lumen.
- 8
While maintaining direct visualization, apply subtle forward pressure. As the sheath tip approaches the mucosa, a small “pillow” or mound of tissue should begin to enter the cystoscopic field of view.
- 9
Advance the needle into the submucosal tissue parallel to the urethral wall.
- 10
Advance to approximately the 1 cm depth marker before injection. After tunneling, the needle tip should lie at the proximal urethra.
- 11
Begin injecting Durasphere using consistent, moderate thumb pressure on the plunger. The flow should be even and smooth, similar to priming the needle.
- 12
Continue injecting until tissue coaptation progresses toward the midline.
- 13
Fully retract the needle. If an opposing injection site is needed, use the opposite InjecSURE needle port to access the opposite side of the urethra.
- 14
Continue injecting until complete circumferential coaptation is achieved. The bladder neck should be closed when viewed with cystoscopic irrigation on.
Periurethral Injection
- 1
Using standard procedure, prepare the patient for cystoscopy.
- 2
Along the periurethral sinus, locate the 2 dimples at the 3 and 9 o’clock position, each within 1 cm of the meatus. Position the needle tip at the dimple with the needle hub parallel to the scope. Note: the proximal half of the needle will be angled away from the scope.
- 3
Penetrate the tissue and advance the needle approximately 3 cm, keeping the needle hub parallel with the scope. The 15° bend guides the tip in an arc to the submucosal tissue between the lamina propria and the muscularis. After advancing, the needle tip should lie in the proximal urethra.
- 4
To verify placement of the needle tip in the submucosal tissue plane, look through the scope while gently wiggling the distal hub of the needle. Only the local area at the needle tip should tent.
- 5
To hydrodissect, inject fluid into the site. Temporary bulking of the tissue should occur. If not seen, reposition more superficially and repeat injection of a small amount of fluid. Switch to the Durasphere syringe when the needle is in the proper position.
- 6
Begin injecting Durasphere using consistent light pressure on the plunger. While injecting, bulking of the tissue should be apparent.
- 7
If circumferential closure of the bladder neck is occurring, continue to inject until complete coaptation is achieved. If the Durasphere product is not flowing circumferentially, continue to inject into the site until the bulge has crossed the midline of the urethra. Reposition the needle on the opposing side and continue injecting. Most procedures will require 4 or more mL of Durasphere.
- 8
Continue injecting until the tissue surrounding the bladder neck coapts. The bladder neck should be closed when viewing with cystoscopic irrigation on.
Post-Procedure Care
Adverse Events and Management
The adverse event data below are drawn from the Durasphere clinical trial, which involved 300 Durasphere treatment injections in 178 patients. There were no deaths among patients injected with Durasphere during the clinical trial. A total of 87.6% of all adverse events were classified as mild, 11.6% as moderate, and 0.8% as severe. The three events classified as severe (chest pain, renal failure, and myocardial infarction) were reported as unrelated to the device or procedure.
For the full adverse event dataset and outcomes data, see the full safety and outcomes data report.
CPT Coding for Cystoscopy with Durasphere Injection
Frequently Asked Questions
INTERESTED IN OUR PRODUCTS?
Talk with one of our trained specialists today for next steps.